How to
Obtain “Correct” Front-Panel Display for
Agilent 33120A Function/Arbitrary Waveform Generator 
The function/arbitrary
waveform generator assumes you are driving a 50-ohm load (this is the power-on
default behavior). If you are driving a high-impedance load (such as the clock
input of a digital gate, or an op-amp input) then you will see that the actual
voltage waveform is two times higher than the displayed value. Follow these
instructions to produce a correct display for high-impedance loads:
Note: Bold face means the
button label, italicized word means button group label
| What to do: | How to do it: | 
| 1. Display the options
  menu | ·       
  Press Shift
  (blue button) ·       
  Press Menu
  Þ On/Off   | 
| 2. Find and select the
  system menu | ·       
  Rotate the
  knob clockwise to select “D: SYS MENU” ·       
  Press Menu
  Þ Ú | 
| 3. Find and select the
  output termination submenu | ·       
  You should
  see “1: OUT TERM” ·       
  Press Menu
  Þ Ú | 
| 4. Select load impedance
  type | ·       
  Rotate knob
  to select “HIGH Z” | 
| 5. Enter your selection
  and return to normal display | ·       
  Press Menu
  Þ Enter |